VIG Model 200/1 FID Methane/Non-Methane (NMHC) Hydrocarbon Analyzer


The VIG Industries, Inc. Model 200/1 is a microprocessor based, oven heated methane / non methane (NMHC) hydrocarbon analyzer for high accuracy, sensitivity and stability. The Model 200/1 measures methane and non methane hydrocarbons at preset intervals. The VIG Model 200/1 uses one(1) Flame Ionization Detector (FID) to obtain methane and non-methane hydrocarbon readings. A sample is fed to the analyzer via an internal heated pump. After the sample reaches the analyzer, a portion of the sample is trapped and fed through a GC (gas chromatography) column to separate the methane from the total and is fed to the FID detector for analysis. As soon as the methane is read, the remaining sample is back flushed through the GC column and fed to the FID detector for analysis of the non-methane component. The LCD display indicates the amount of methane and non-methane hydrocarbons.
The Model 200/1 utilizes one flame ionization detector (FID). The detector, all sample lines, GC column and all solenoid valves that come in contact with the sample are maintained in a controlled, heated oven to prevent condensation, and to provide reliable performance in the analysis of concentrations of wide varieties of hydrocarbons in gaseous mixtures and in ambient air.
Standard Specifications
Analysis Method: Oven Heated, Flame Ionization Detector (FID)
Separation Method: Column
Precision: +/- 1% of Full Scale
Response Time:
Methane - Approximately 40 seconds, updated every 3 minutes
Non methane - Approximately 70 seconds, updated every 3 minutes
Zero Drift: +/- 1% of Full Scale Per 24 Hours
Span Drift: +/- 1% of Full Scale Per 24 Hours
Linearity: Within 1% of full scale
Oxygen Synergism : Within 1% of full scale within selected range
Ranges: (4 Ranges per amplifier, two amplifiers per analyzer, one amplifier for total hydrocarbons, one amplifier for methane and non-methane)
Standard Ranges: 0-100, 1000, 10000, 100000ppm (LDL 0.1ppm), 0-10, 100, 1000, 10000ppm (LDL 0.01ppm)
Display: Graphic, backlit, 240W x 64H pixels, high contrast, wide viewing angle
Outputs: 0-1, 5, or 10 Volts & optional 4-20mA
Alarms: Flame Out & Concentration, Latching
Safety: Flame Out Indicator with Fuel Shut-Off System Oven Temperature
Operating Temperature: 275°F, Adjustable 200 to 300° F
Sample Pump: Heated 4 Liters/Minute standard or Heated 14 Liters/Minute optional
Ignition: Automatic (can be set to manual by operator)
Glow Plug: Main and spare glow plug, selectable by switch on back panel
Dimensions: 19" front panel, 16.75" wide chassis, 18" deep chassis, 21" deep w/ fittings and handles, 9" high
Weight: 55 to 65 pounds depending on options
Operating Requirements
Fuel: Hydrogen (UHP Grade) @ 18psi incoming pressure
Combustion Air: Zero Grade Air @ 18PSI
Operation Air: Compressed Air (Instrument Air - Dry / Oil-Less) @ 60PSI
Zero Gas: Zero Air (UHP Grade) @ 10PSI
Calibration Gas: User Selected @ 10PSI
Carrier Gas: Nitrogen (UHP Grade) @ 30PSI
Power: 115Vac, 60Hz or optional 220Vac, 50Hz
Ambient Temperature: 32 to 110º F