Rental/Sales
800 399-4CEM
(800 399-4236)

rfqrfqTeledyne-API Model 300E
Gas Filter Correlation CO Analyzer




The Model 300E measures low ranges of carbon monoxide by comparing infrared energy absorbed by a sample to that absorbed by a reference gas according to the Beer-Lambert law. This is accomplished with a Gas Filter wheel which alternately allows a high energy light source to pass through a CO filled chamber and a chamber with no CO present. The light path then travels through the sample cell, which has a folded path of 14 meters.

The energy loss through the sample cell is compared with the span reference signal provided by the gas filter to produce a signal proportional to concentration, with little effect from interfering gases within the sample. This design produces excellent zero and span stability and a high signal to noise ratio allowing extreme sensitivity. Multi-tasking software gives real time indication of numerous operating parameters and provides automatic alarms if diagnostic limits are exceeded.

Built-in data acquisition and internal memory allows logging multiple parameters including average and instantaneous values, calibration data and operating parameters. Stored data are easily retrieved through the serial port or optional Ethernet port via our APIcom software or from the front panel, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends.

Specifications
Ranges: 0-1 ppm to 0-1,000 ppm full scale, user selectable; Dual ranges and auto ranging supported
Units: ppb, ppm, µg/m3, mg/m3
Zero noise: < 0.02 ppm
Span noise: < 0.5% of reading above 5 ppm (RMS)
Lower Detectable Limit (LDL): 0.04 ppm
Zero drift: < 0.1 ppm/24 hours, 0.2 ppm/7 days
Span drift: < 0.5% of reading/24 hours, 1% of reading/7 days
Lag time: 10 seconds
Rise and Fall Time: < 60 seconds to 95%
Linearity: 1% of full scale
Precision: 0.5% of reading
Sample Flow Rate: 800 cm3/min ±10%
Operating Temperature Range: 5 - 40° C (with EPA equivalency)
Dimensions (HxWxD): 7” (178 mm) x 17” (432 mm) x 23.5” (597 mm)
Weight: 40 lbs (18 kg)
Power: 100V - 120V, 220V - 240V, 50/60 Hz, 250W
Analog outputs: 10V, 5V, 1V, 0.1V, selectable
Recorder Offset: ±10%
Serial Outputs: Serial Port 1: RS-232 (DB-9M)
Serial Port 2: standard RS-232 or optional RS-485 (DB-9F), Ethernet
Status (Digital): 8 outputs, 6 inputs (opto-isolated), 4 alarm outputs (optional)
Current Output: Optional 4-20mA, select up to three channels
Approvals: USEPA RFCA-1093-093,
MCERTS certified Sira MC050069/00
EN14626 Approved, and others.